EFFLUENT TREATMENT

EFFLUENT TREATMENT

BIOCONTAMINATED EFFLUENT TREATMENT

The aim of EFFLUENT TREATMENT SYSTEM is to execute a thermal sterilization at 134 °C for all potentially contaminated effluents generated by the establishment, in safety conditions, both for the staff responsible for such function and for the environment.

TREATMENT

The system consists of one or more digesters where appropriate sterilization of effluents collected takes place. Subsequently, these effluents are cooled and discharged into the sewer under sterile conditions.

Additionally, all the circuits involved in the processing system are sterilized automatically, each time a risk treatment is performed.

Common remarks:

For proper operation of the system, the following supplies should be at disposal:

  • Steam
  • Compressed air (a dual/redundant compressor is recommended)
  • Softened water

50 LITERS TREATMENT CAPACITY/PROCESS

Effluent treatment system, comprising:

  • Digester with jacket, 50l treatment capacity per sterilization process
  • With self-cleaning circuit at the end of the sterilization cycle
  • Effluent cooling prior to the discharge to the drain
  • Accumulator for total volume of 1200 liters, suitable for disinfection by steam sterilization cycle
  • Control system with industrial Siemens PLC, touchscreen and process chart recorder

400 LITERS TREATMENT CAPACITY/PROCESS

Effluent treatment system, comprising:

  • Digester with jacket, 400l treatment capacity per sterilization process
  • With self-cleaning circuit at the end of the sterilization cycle
  • Effluent cooling prior to the discharge to the drain
  • Accumulator for total volume of 1500 liters, suitable for disinfection by steam sterilization cycle
  • Control system with industrial Siemens PLC, touchscreen and process chart recorder

800 LITERS TREATMENT CAPACITY/PROCESS

Effluent treatment system, comprising:

  • 2 digesters with jacket, 800l treatment capacity per sterilization process (operating in parallel)
  • With self-cleaning circuit at the end of the sterilization cycle
  • Effluent cooling prior to the discharge to the drain
  • Accumulator for total volume of 1500 liters, suitable for disinfection by steam sterilization cycle
  • Control system with industrial Siemens PLC, touchscreen and process chart recorder

3X650 LITERS TREATMENT CAPACITY/PROCESS

Effluent treatment system, comprising:

  • 3 digesters with jacket, 650l treatment capacity per unit. The 3 units are able to operate simultaneously, if necessary
  • With self-cleaning circuit at the end of the sterilization cycle
  • Effluent cooling prior to the discharge to the drain
  • Control system with industrial Siemens PLC, touchscreen and process chart recorder

3X2000 LITERS TREATMENT CAPACITY/PROCESS

Effluent treatment system, comprising:

  • 3 digesters with jacket, 2000l treatment capacity per unit. The 3 units could operate simultaneously, if necessary
  • With self-cleaning circuit at the end of the sterilization cycle
  • Effluent cooling prior to the discharge to the drain
  • Control system with industrial Siemens PLC, touchscreen and process chart recorder